Assaying multiple 125 I seeds with the well-ionization chamber SourceCheck 4π 33005 and a new insert

2015

PURPOSE To provide a practical solution that can be adopted in clinical routine to fulfill the AAPM-ESTRO recommendations regarding quality assurance of seeds used in prostate permanent brachytherapy. The aim is to design a new insert for the well-ionization chamber SourceCheck(4π) 33005 (PTW, Germany) that allows evaluating the mean air-kerma strength of up to ten (125)I seeds with one single measurement instead of measuring each seed individually. MATERIAL AND METHODS The material required is: a) the SourceCheck(4π) 33005 well-ionization chamber provided with a PTW insert to measure the air-kerma strength S K of one single seed at a time; b) a newly designed insert that accommodates ten s…

Original Paperprostatebusiness.industrybrachytherapySingle measurementMean valueAnalytical chemistryfood and beveragesinsertquality assuranceseedsClinical routinecomputer.software_genreInsert (molecular biology)Calibration coefficientOncologyIonization chamberMedicineRadiology Nuclear Medicine and imagingData miningbusinesscomputerwell chamberJournal of Contemporary Brachytherapy